Tesla blood pumps demonstrate a reduced propensity for hemolysis and thrombosis compared with vane blood pumps. Considering the restricted driving force within the secondary flow channel of vane blood pumps, along with the low hydraulic efficiency of conventional Tesla blood pumps and their internal flow characteristics that significantly contribute to hemolysis and thrombosis, this study introduces a set of vanes atop the rotor of the Tesla blood pump. This forms a dual-fluid domain rotor, and an axial dual-outlet volute shell structure is adopted to realize the separation of the fluid domains. Through numerical simulations of the new structure, a comparative analysis was conducted in this study on the internal flow characteristics of double-outlet and single-outlet volute shells, and symmetric and asymmetric cross-sections of the same rotor. The results indicate that the flow field distribution is more uniform under the double-outlet volute shell structure, and overall energy dissipation is decreased. After implementing the double-outlet design, in the asymmetric cross-section, compared with the symmetric cross-section, the fluid velocity gradient and turbulent kinetic energy at the tongue of the septum are reduced, and the fluid velocity gradient at the convergence of the diffuser tube outlets are also decreased. The maximum scalar stress is lower, and the decline in head and efficiency is mitigated. Moreover, compared with the single-outlet volute shell, the hemolysis index in the asymmetric cross-section is reduced. In summary, this paper proposes a novel dual-outlet centrifugal disk blood pumps, which can provide a reference for the structural design and performance optimization of magnetically levitated centrifugal blood pumps.